|
Groundwater Professional Bulletin Board
POSTED: 04/30/2003
ADEQUATE SAMPLING DURING OVER EXCAVATION
Over-excavation can be considered expedited corrective action, as described under
Iowa Administrative Rule 567 IAC 135.12 (455B), and, as such, may be conducted at any
time during the RBCA process as provided therein.
Field screening should be conducted before, during, and after excavation to verify
removal of the soil source has been completed. Soil samples must be obtained from the
completed excavation for field screening and laboratory analysis to confirm remaining
soil contaminant concentrations are below Tier 1 levels. The sampling guidelines
described in paragraph IAC 135.12(11)"c" and Tier 2 Guidance section 5.6
apply to the finished excavation limits. Adequate documentation of vapor screening
(PID, FID, or similar), visual, and/or olfactory observations, field screening, and
sample selection justification must be provided in the excavation report. Samples for
laboratory analysis shall be collected from not more than one foot into the base and
sidewalls of the excavated area.
Field Screening. As described under paragraph IAC 135.12(11)"c"
and Tier 2 Guidance section 5.6, adequate field screening shall consist
of at least 1 (one) field screening sample taken and reported from each
100 square foot area (or portion thereof) of the base and each 100 square foot
area (or portion thereof) of each sidewall of the completed excavation. In some cases
the area of a sidewall or the base may be less than 100 ft2. In these cases,
at a minimum, at least 1 (one) field screening sample must be taken and reported from
each sidewall and the base of the completed excavation for a total of
5 (five) field screening samples. Sample locations should be selected for field
screening based on visual or olfactory indications of possible contamination. All
observations and vapor screening results must be documented in the excavation
report. In some cases, based on field observations (discolored soils, petroleum odors,
etc), more than 1 (one) field screening sample from each 100-ft2 area may
be warranted.
Laboratory Samples. As described under paragraph IAC 135.12(11)"c"
and Tier 2 Guidance section 5.6, adequate sampling for laboratory analysis shall
consist of 1 (one) lab sample taken from each 400 square foot area (or portion thereof)
of the base and each 400 square foot area (or portion thereof) of each
sidewall of the completed excavation. In some cases the area of a sidewall or the base
may be less than 400 ft2. In these cases, at a minimum, at least 1 (one)
soil sample must be submitted for analysis from each sidewall and the base of
the completed excavation. Samples to be submitted for laboratory analysis will be
selected, based on highest PID, from the field screening samples within the designated
area. In some cases, based on PID readings, more than 1 (one) lab sample from each
400-ft2 area may be warranted.
In the event excavation is conducted in such a manner as to create more than four
sidewalls, field screening and laboratory samples must be obtained from each sidewall.
Example 1: Given excavation limits of 5 feet deep x 6 feet wide x 10 feet
long. The minimum number of field screening samples = 1 (one) for each sidewall and
1 (one) for the floor of the excavation, totaling 5 (five) field screening samples.
The minimum number of samples submitted for lab analysis would be 5 (five); 1 (one)
from each sidewall and 1 (one) from the base of the excavation.
Example 2: Given excavation limits of 10 feet deep x 10 feet wide x 35 feet
long. The minimum number of field screening samples = 1 (one) for each sidewall along
the width of the excavation, 4 (four) for each sidewall along the length of the
excavation, and 4 (four) for the floor of the excavation, totaling 14 field screening
samples. The minimum number of samples submitted for lab analysis would be 5(five);
1 (one) from each sidewall and 1 (one) from the base of the excavation.
Example 3: Given excavation limits of 12 feet deep x 38 feet wide x 65 feet
long. The minimum number of field screening samples = 5 (five) for each sidewall along
the width of the excavation, 8 (eight) for each sidewall along the length of the
excavation, and 25 for the floor of the excavation, totaling 51 field screening samples.
The minimum number of samples submitted for lab analysis would be 15; 2 (two)
from each sidewall and 7 (seven) from the base of the excavation. NOTE: In this
instance, selecting 7 lab samples based on highest PID does not refer to selecting the
highest 7 PID samples from the entire excavation bottom. Instead, this refers to
selecting the highest PID sample from each of seven 400 ft2 areas scattered
across the base of the excavation.
Finally, the Tier 2 must be re-evaluated and all the affected portions of the
Tier 2 Report, including the Tier 2 disk, must be submitted with the Over Excavation
Report. The over excavation analytical results must be entered in the Tier 2 portion
of the software for proper re-evaluation of the soil pathways. The results of all
prior soil sample locations, removed during the course of the over excavation,
must be marked as `ignore' in the soil data adjustment section. All sample information
(concentrations, location, elevation, etc.) must be entered in the data table.
Contact
Elaine Douskey at (515)281-8011 or
elaine.douskey@dnr.state.ia.us
if you have questions.
|